Technical Parameters |
|
|
As per Standards |
BS EN 12895:2015 |
|
Magnetic Field Strength |
1 A/m – 1000 A/m |
|
Current Distortion Rate |
<8% |
|
Pulse Interval |
1 s - 9,999 s |
|
Test Duration |
1 A/m ~ 100 A/m: continuous from 1 s to 28,800 s; ≥100 A/m: continuous for 10 s |
|
Triggering Mode |
Automatic, manual or external signal triggering |
|
Coil Dimension |
3,000*3,000 mm |
|
Coil Shape |
Rectangle |
|
Test Direction |
90°/180° |
|
Coil Turns |
60-turn |
|
General Parameter |
|
|
Display screen |
5.7” TFT touch screen |
|
Operating power supply |
110 V / 220 V, ±10%, 50/60 Hz (AC 220 V 50 Hz by default) |
|
Max. Power consumption |
200 W |
|
External Trigger Input |
BNC, 5 V TTL |
|
Running Control Input |
BNC, 5 V TTL |
|
Warning Light Output |
0-220 V AC 10 A |
|
Safety Circuit |
Work will be stopped when safety circuit is open |
|
Failure Detection |
The failure information will be displayed on LCD in front panel when failure occurs and interrupt the working of instrument. |
|
Working Status Indication |
LED indication and LCD display on front panel |
|
Earthing Mode |
Flat earth line |
|
Working power |
380 VAC/32 A |
|
Dimension |
22U |
|
Weight |
Approx.130 kg |
|
Operating temperature |
15 ℃~ 35 ℃ |
|
Operating humidity |
45%–75% |
|
Operating air pressure |
86 kPa – 106 kPa |
|
Accessories |
|
User Manual, aviation connector test line, 3-core power line, test line, earth line, fuses (spare parts) |
|
Options |
|
Magnetic field coil TCXS 3360 |