Technical Parameters |
|
Test Voltage |
90 V ~ 2500 V , ±10% |
Waveform 1 |
|
Calibration Voltage |
2400 V , ±10% |
Rise Time (0% ~ 100%) |
≤35 μs |
Pulse Width (0% ~ 50%) |
≥63 μs |
Pulse Width (0% ~ 0%) |
≥86 μs |
Waveform 2 |
|
Calibration Voltage |
1000 V , ±10% |
Rise Time (0% ~ 100%) |
≤62 μs |
Pulse Width (0% ~ 50%) |
≥119 μs |
Pulse Width (0% ~ 0%) |
≥178 μs |
Waveform 3 |
|
Calibration Voltage |
500 V , ±10% |
Rise Time (0% ~ 100%) |
≤60 μs |
Pulse Width (0% ~ 50%) |
≥130 μs |
Pulse Width (0% ~ 0%) |
≥280 μs |
Waveform 4 |
|
Calibration Voltage |
100 V , ±10% |
Rise Time (0% ~ 100%) |
≤65 μs |
Pulse Width (0% ~ 50%) |
≥400 μs |
Pulse Width (0% ~ 0%) |
≥1250 μs |
Calibration Load |
200 Ω±5% series 0.1 μF±10% |
Polarity |
Positive, negative, alternative |
Test Interval Time |
3 s ~ 999 s |
Counter |
1 ~ 10000 |
General Parameters |
|
Display Screen |
5.7-inch TFT touch screen |
Working Power |
AC 220 V ± 10%,50/60 Hz |
Fuses |
6A |
Maximum Power Consumption |
200 W |
Memory Space |
Unlimited (PC) |
Communication Methods |
Ethernet LAN, RJ45 |
Trigger Mode |
Auto, manual, external trigger |
External Trigger Input |
BNC , 5V TTL, normally open node, closed trigger |
CRO Trigger Output |
BNC , 5V TTL |
Operation Control Input |
BNC , 5V TTL |
Failure Detection |
When it fails, the front panel LCD displays and the instrument stops working |
Working Status Indication |
Front panel LED indication, LCD display |
Grounding Connection Method |
Use a flat grounding wire |
Dimension |
19 inches/4U 450 mm(W)*190 mm(H)*620 mm(D) |
Weight |
Approx. 22 kg |
Ambient Temperature |
15℃ ~ 35℃ |
Relative Humidity |
45%~75%,RH (no condensation) |
Atmospheric Pressure |
86 kPa ~ 106 kPa |