Introduction
CCM 0301M card broadband current monitoring clamp is used for measuring alternating current and time-domain pulse current from 10 Hz to 1 MHz. By electromagnetic coupling, the time harmonic signal current or pulse signal current is transmitted to the recording device in the form of voltage, which has the characteristics of waveform distortion free and physical isolation from the current carrier. Meet the measurement requirements of various standards for harmonic signals of 1 MHz and below, and transient pulse signals with a rise time greater than 350 ns. Widely used in electromagnetic compatibility conducted sensitivity (immunity) testing.
Features
> Meet the measurement requirements for harmonic signals of 1 MHz and below
Technical Parameters |
|
Sensitivity* |
0.001 V/A |
50 Ω |
|
Maximum Peak Current |
20,000 A |
Maximum Continuous Current(effective value) |
600 A |
Rise Time |
350 ns |
An Second Accumulation |
12 A•s |
Lower Limit Frequency (-3 dB) |
10 Hz |
Upper Limit Frequency (-3 dB) |
1 MHz |
Inner Diameter |
45 mm |
Outer Dimensions |
151 mm(W)*45 mm(H)*160 mm(D) |
Interface Type |
BNC type |
Operation Temperature |
|
Weigh |
Approx. 1.7 kg |
*When the broadband current monitoring clamp is connected to 50 Ω, the sensitivity (transfer impedance) is 0.0005 V/A. |