Features:
> 5.7-inch color touch screen front panel operation;
Naming Rules:
Note: External network and other accessories have independent models and are not reflected in the host model; -300-20/32 represents the presence of a built-in coupling decoupling network.
General Parameters |
|
Display |
5.7-inch TFT touch screen |
Scope of Working Power Supply |
100 ~ 264 V AC, 47 ~ 63 Hz |
Fuse |
6:00 AM |
Maximum Power Consumption |
300 W |
Communication Method |
Ethernet LAN |
External Control Mode |
25 needle parallel thread |
External Trigger Input |
BNC, 5 V TTL |
CRO Triggers Output |
BNC, 5 V TTL |
Operation Control Input |
BNC, 5 V TTL |
Pulse Triggering Method |
Manual, automatic, externally triggered |
External Synchronous Input |
20 ~ 400 V, 45 ~ 65 Hz |
Warning Light Output |
Multi core connector output, matched with external alarm light module (optional) |
Safety Circuit |
Short circuit of safety loop, stop working when the safety loop is open circuit |
Failure Detection |
When it fails, the front panel LCD displays and interrupts the instrument's operation |
Instrument Working Status Indication |
LED indication, LCD display |
Instrument Grounding Connection Method |
Use a flat grounding wire |
Chassis Size |
6U L * W * H: 620 * 450 * 280 mm |
Instrument Weight |
About 50 kg |
Environmental Scope |
15℃ ~ 35℃ |
Pressure Range |
86 kPa ~ 106 kPa |
Humidity Range |
45% ~ 75% |
Built in coupling/decoupling network (single-phase fully automatic) |
|
EUT Carrying Capacity |
AC 300 V 20 A/32 A (optional) 50/60 Hz |
DC 300 V 20 A/32 A (optional) |
|
EUT Power Input and Output |
4mm banana plug cable |
EUT Voltage Monitoring Output |
BNC output, 100 V: 1 V |
EUT Current Monitoring Output |
BNC output, 10 A: 1 V |
Synchronization Method |
Internal synchronization, external synchronization, asynchronous |
Internal Synchronization |
0°~360°, 1° step setting or random mode |
1.2/50 μs Combined Wave Coupling/decoupling |
Built in single-phase automatic coupling/decoupling network |
Note: The CWS 600 (S) configuration only has a source, while the CWS 600 (S-300-20) comes with a built-in coupling decoupling network |
|
IEC-61000-4-5 Surge Immunity Test |
|
Test Voltage |
0.25 kV ~ 6 kV ±10% |
Test Current |
0.125 kA ~ 3 kA ± 10% |
Voltage Waveform |
1.2 μs ± 30% , 50 μs ± 20% |
Current Waveform |
8 μs ± 20%, 20 μs ± 20% |
Output Impedance |
2 Ω, 12 Ω |
Test Interval Time |
6 ~ 99 s(the shortest depends on the test voltage) |
Number of Experiments |
1~999 times |
Experimental Mode |
Optional scheduling mode |
Polarity |
Positive, negative, alternate |
Calibration Capacitor |
18 μF built-in |
Coupling Resistance |
0 Ω, 10 Ω |
Coupling Capacitor |
Built in at 9 μF and 18 μF |
Surge Voltage Peak Detection |
LCD display, BNC output 1000 V: 1 V |
Surge Current Peak Detection |
LCD display, BNC output 500 A: 1 V |
IEC-61000-4-5 Communication Wave Test |
|
Test Voltage |
0.25 kV ~ 6 kV (±10%) |
Test Current |
6.25 A ~ 150 A ±10% |
Voltage Waveform |
10 μs ± 30% , 700 μs ± 20% |
Current Waveform |
5 μs ± 20%, 320 μs ± 20% |
Output Impedance |
15 Ω, 40 Ω |
Test Interval Time |
6 ~ 99 s(the shortest depends on the test voltage) |
Number of Experiments |
1~999 times |
Experimental Mode |
Optional scheduling mode |
Polarity |
Positive, negative, positive/negative |
Note: CWS 600 (T) should be selected for communication waves |
CWS 600 Selection Guide List |
|||
Generator |
1.2/50 μs |
10/700 μs |
CDN single-phase three wire network |
Compact immunity test system CWS 600(ST-300-20) |
√ |
√ |
√ |
Compact immunity test system CWS 600(S-300-20) |
√ |
|
√ |
Compact immunity test system CWS 600(S) |
√ |
|
|
Compact immunity test system CWS 600(T) |
|
√ |
|
Optional module |
|||
Lightning surge coupling/Decoupling network SPN xxxxT10 series |
√ |
|
Can be customized according to EUT voltage and current level requirements |
Non shielded symmetrical high-speed communication line lightning surge coupling decoupling network CDN 405T8A1 |
√ |
√ |
|
Non shielded symmetrical high-speed communication line lightning surge coupling decoupling network CDN 405T8-C |
√ |
|
|
List of Testing and Measurement Selection Guidelines
Instrument
name and model
1.2/50 μs
10/700 μs
Differential
probe VCF-80
√
√
CM 0220M
√
√
10 kV surge calibration capacitor module
CCM 1000
√
√
Immunity testing software
√
√
Broadband current monitoring clamp
CoreLab