General Parameters |
|
|
Display |
5.7-inch TFT touch screen |
|
Scope of Working Power Supply |
100 ~ 264 V AC, 47 ~ 63 Hz |
|
Fuse |
6A |
|
Maximum Power Consumption |
300 W |
|
Communication Method |
Ethernet LAN |
|
External Control Mode |
25 needle parallel thread |
|
External Trigger Input |
BNC, 5V TTL |
|
CRO Triggers Output |
BNC, 5V TTL |
|
Operation Control Input |
BNC, 5V TTL |
|
Pulse Triggering Method |
Manual, automatic, externally triggered |
|
External Synchronous Input |
20 ~ 400 V, 45 ~ 65 Hz |
|
Warning Light Output |
Multi core connector output, matched with external alarm light module (optional) |
|
Safety Circuit |
Short circuit of safety loop, stop working when the safety loop is open circuit |
|
Failure Detection |
When it fails, the front panel LCD displays and interrupts the instrument's operation |
|
Instrument Working Status Indication |
LED indication, LCD display |
|
Instrument Grounding Connection Method |
Use a flat grounding wire |
|
Chassis Size |
6U L * W * H: 620 * 450 * 280 mm |
|
Instrument Weight |
About 50 kg |
|
Environmental Scope |
15℃ ~ 35℃ |
|
Pressure Range |
86 kPa ~ 106 kPa |
|
Humidity Range |
45% ~ 75% |
|
IEC-61000-4-9 Pulse Magnetic Field Test |
|
|
Magnetic Field Strength (1*1 m coil) |
100 A/m ~ 1200 A/m |
|
magnetic Field Strength (1*2.6 m coil) |
100 A/m ~ 880 A/m |
|
Coil Waveform (1*1 m Single Turn) |
8 μs(+ 2.4 μs /- 0.8 μs),20 μs(+ 6 μs/-2 μs) |
|
Coil Waveform (1*2.6 m Single Turn) |
8 μs(+ 3.2 μs/- 0.8 μs),20 μs( + 8 μs/ -2 μs) |
|
Test Interval Time |
5 ~ 99 s(the shortest depends on the magnetic field strength) |
|
Number of Experiments |
1~999 times |
|
Experimental Mode |
Optional scheduling mode |
|
Polarity |
Positive, negative, first positive and then negative |
|
Magnetic Field Coil Size |
1 m x 1 m, other |
|
Shape of Magnetic Field Coil |
Rectangle, other |
|
Standard Accessories |
||
|
Power Cord, Fuse, Test Wire, Grounding Wire, Instruction Manual |
||
|
Optional Accessories |
||
|
Magnetic Field Coil |
Model: TCXS series |
|
|
Testing and Measurement Selection Guide List |
||
|
Broadband Current Monitoring Clamp |
Model: CM 0220M |
|
|
Immunity Testing Software |
Model: CoreLab |
|