Technical Parameter |
|
|
Test Current Range |
20 A ~ 6000 A(±10% ±1 A) |
|
Current Wavefront Time |
8 μs ± 10%, wavefront time definition: 1.25 * (10%~90%) |
|
Half Peak Time of Current |
20 μs ± 10%, definition of half peak time: 1.18 * (50%~50%) |
|
Equivalent Output Impedance |
≤14.5 Ω(20 A≤I<200 A) |
|
Output Polarity |
Positive, negative, first positive and then negative |
|
Test Interval |
5 s~180 s (the shortest time depends on the test voltage) |
|
Test Frequency |
1 ~ 10000 |
|
Pulse Triggering Method |
Manual, automatic, externally triggered |
|
General Parameters |
|
|
Display |
5.7-inch TFT touch screen |
|
Scope of Working Power Supply |
AC 220 V, ± 10%, 50 Hz ± 5% |
|
Fuse |
6A |
|
Maximum Power Consumption |
300 W |
|
Communication Method |
Ethernet LAN |
|
External Trigger Input |
BNC, 5V TTL |
|
CRO Triggers Output |
BNC, 5V TTL |
|
Operation Control Input |
BNC, 5V TTL |
|
Instrument Grounding Connection Method |
Use a flat grounding wire |
|
Chassis Size |
6U L * W * H: 620 * 450 * 280 mm |
|
Instrument Weight |
About 29 kg |
|
Environmental Scope |
15 ℃ ~ 35℃ |
|
Pressure Range |
86 kPa ~ 106 kPa |
|
Humidity Range |
45% ~ 75% |
|
Standard Accessories |
|
|
Test Line, Power Line, Grounding Wire, Fuse (spare parts), Instruction Manual, Factory Report. |
|
|
Optional Accessories |
|
|
CPN 2216S current wave coupling decoupling network |
The maximum coupled pulse current can reach 6 kA. EUT maximum load capacity AC 220 V 16 A single-phase DC 220 V 16 A |
Tel: +86 (0)512 6807 7192