Technical Parameters |
|
|
Rated Voltage |
1 V-50 V,0.1 V step |
|
Voltage Dips / Surge Voltage |
0-120% |
|
Voltage Dips / Surge Duration |
0.1 ms-59.999 s |
|
Voltage Rise / Fall Time |
< 50 μs (generator loaded with 100 Ω resistive load) |
|
Overshoot/ undershoot of the Output Voltage |
Less than 10 % of the change in voltage (generator loaded with 100 Ω resistive load) |
|
Time Intervals |
0.10.1 s – 60 s (the min. time depends on the test voltage) |
|
Measuring Accuracy |
Voltage:1% (±0.1 V); Current: 1% (±0.1 A); |
|
Max. Output Current (Steady State) |
≤ 5 A |
|
Number of Tests |
1 – 9,999 |
|
Ripple Content |
Less than 1% of the output voltage |
|
Load Adjustment Rate |
Output voltage variation with the load (0 to rated current) <5% |
|
Output Impedance during a Short Interruption |
>100 kΩ |
|
General Parameters |
|
|
Display Screen |
5.7” TFT touch screen |
|
Working Power Supply |
AC 110 V / 220 V (±10%), 50 Hz / 60 Hz (Default AC 220 V 50 Hz in mainland China |
|
Fuse |
10 A |
|
Max. Power Consumption |
1,000 W |
|
User’s Memory Space |
Infinite (PC) |
|
Communication Mode |
Ethernet LAN, RJ45 |
|
Working Status Indication |
LED indication and LCD display on front panel |
|
Dimension |
19”, 4 U x 2 pcs |
|
Weight |
Approx. 40 kg |
|
Ambient Temperature |
15℃ - 35℃ |
|
Relative Humidity |
45%–75% |
|
Atmospheric Pressure |
86 kPa–106 kPa |