| Technical Parameters | ||||
| Model No. | SPN 3820T20 | SPN 3820T30 | ||
| Testing Standard | GB/T 17626.5, IEC 61000-4-5 | GB/T 17626.5, IEC 61000-4-5 | ||
| EUT Capacity | AC three-phase-5-wire 380 V(L1,L2、L3、N、PE)access ; DC-1 220 V (L1,N) access Current 20 A, 30 A, 50 A and 100 A are optional | AC three-phase-5-wire 380 V(L1,L2、L3、N、PE) access; DC 200 V (L1,N) access Current 20 A, 30 A, 50 A and 100 A are optional | ||
| Capacitive Coupling | ||||
| Testing Waveform | Combination Wave(1.2/50 μs & 8/20 μs) | Combination Wave(1.2/50 μs & 8/20 μs) | ||
| Surge Voltage Injection | 0.5 kV -20 kV | 0.5 kV -30 kV | ||
| Coupling Surge Current | 0.25 kA -10 kA | 0.25 kA -15 kA | ||
| Coupling Capacitance | Common-mode:9 μF(+10 Ω) Differential-mode:18 μF | Common-mode:9 μF(+10 Ω) Differential-mode:18 μF | ||
| Clearance Coupling | ||||
| Testing Waveform | 8/20 μs、10/350 μs、10/1000 μs etc | |||
| Surge Voltage Injection | 2 kV -20 kV | 2 kV -30 kV | ||
| Coupling Surge Current | 2 kA -20 kA | 2 kA -30 kA | ||
| Coupling Capacitance | Gas Discharge Tube(100 kA) | Gas Discharge Tube(100 kA) | ||
| Decoupling Inductance | 1.5 mH (optional 20mH can meet the requirement of GB/T 3483) | |||
| Mains Switching Mode | AUTO/MANUAL | |||
| Coupling Superposition Mode | AUTO/MANUAL | |||
| General Parameters | ||||
| Working Mains | AC 110 V/220 V(±10%),50 Hz /60 Hz(AC 220V in Mainland China) | |||
| Ambient Temperature | 15℃-35℃ | |||
| Relative Humidity | 45%-75% | |||
| Atmospheric Pressure | 86 kPa-106 kPa | |||
| Cabinet Dimension | 19 Inches / 22 U(600 mm x800 mm x1250 mm) / 35 U(600 mm x800 mm x1820 mm) | |||
| Weight | Approx. 150 kg | |||