Magnetic coil with a handle FESP 5132 can product standard magnetic field strength at 150 kHz voice frequency. It is mainly used for magnetic field immunity test as per EN 55103, MIL-STD-461 E and ISO 11452-8. The output magnetic field can be up to 1500A/m for short duration. The magnetic field is proportional to coil current. The distance of coil and EUT surface can be 50mm.
Technical parameters |
|
Diameter |
120 mm |
Turns |
20 |
Wire diameter |
2 mm copper (AWG 12) |
Max. Coil current |
20A, 5 min |
Nominal coil current |
10A continuous |
Max. Magnetic field strength |
1500 A/m, 5 min |
Nominal magnetic field strength |
750 A/m continuous |
Magnetic field strength, 1A coil current(distance to loop: 50mm) |
75.56 A/m (95 µT) |
Current for 1 A/m:(distance to loop: 50mm) |
13.23 mA |
Center of empty coil to measurement plane |
50 mm |
Mechanical dimension |
0.16 m x 0.25 (0.3) m |
Connection terminal |
4 mm banana plug |
Frequency range |
15 - 150 kHz |
Inductance |
80 µH |
Resistance |
0.3 Ω |
Weight |
1.2 kg |
Accessories |
||
Connection cable |
2 |
|
Calibration report |
1 |
|
User manual |
1 |
|