10/160 Combination Wave Parameter |
|
|
Output voltage |
0.25~1.65 Kv ± 5% |
|
Output current |
31~220 A |
|
Output voltage wave |
6~10/160~260 μs |
|
Output current wave |
5~10/160~210 μs |
|
10/560 Combination Wave Parameter |
|
|
Output voltage |
0.25~0.88 kV ± 5% |
|
Output current |
33.3~115 A |
|
Output voltage wave |
6~10/560~860 μs |
|
Output current wave |
5~10/560~760 μs |
|
9/720 Combination Wave Parameter |
|
|
Output voltage |
0.25~1.65 kV ± 5% |
|
Output current |
6.25~41.25 A |
|
Output voltage wave |
9±2.7/720±144 μs |
|
Output current wave |
5±1.5/320±64 μs |
|
2/10 Combination Wave Parameter |
|
|
Output voltage |
0.25~2.75 kV ± 5% |
|
Output current |
100~1100 A |
|
Output voltage wave |
1~2/10~19 μs |
|
Output current wave |
1~2/10~19 μs |
|
Single Phase CDN Parameters |
|
|
Decoupling waveform |
2/10 μs |
|
Decoupling amplitude |
Less than 6 kV / 3 kA |
|
Coupling capacitance |
18 μF |
|
Decoupling inductance |
1.5 mH |
|
Coupling mains |
Single phase AC 220 V 50 Hz, 20 A |
|
Test mode |
Automatic switch |
|
General Parameter |
|
|
Output polarity |
+/- |
|
Number of discharging |
1~999 |
|
Interval time |
30~999 s |
|
Trigger mode |
Automatic or manual, single trigger available |
|
Current measurement |
Internal 0.005 V / A current sensor |
|
Power supply |
AC 110 V / 220 V, ±10%, 50/60 Hz (default AC 220 V) |
|
Temperature |
15~35 ℃ |
|
Humidity |
45%-75% |
|
Ground resistance |
No more than 0.5 Ω |
|
Dimension |
19” /22U 600x800x1250 mm |
|
Weight |
Approx. 130 kg |
|
Optional Accessories |
|
|
Oscilloscope |
Tektronix MDO series |
|
Wide-band Current Monitor |
CM 0220M |
|
Voltage probe |
HILOK VCF-80, 8 kV differential |
|
Signal line CDN |
CDN |
|
Test box |
PA350 |