The PFS 69200T1F simulates the impact of voltage dips, interruptions, and voltage variations on electrical and electronic equipment in the power grid. The reasons for these voltage changes are due to faults in the power grid, substation equipment, or sudden large changes in load. Testing such phenomena is to verify whether electrical and electronic equipment can avoid unsafe operating conditions when encountering voltage fluctuations. The device is equipped with controllable semiconductor switches, which can withstand instantaneous impulse currents more than 1000 A and have electronic short circuit protection function and can fully be compliant with IEC 61000-4-34 and GB/T 17626.34.
> 5.7 inch color touch screen front panel operation
> Fully compatible with IEC 61000-4-34 testing
> Voltage/current measurement
> Voltage limiter, current limiter, overvoltage, overcurrent protection function
> Programmable test procedures
> Ethernet RJ45 interface, used for PC remote control and printing test reports
|
General parameters |
|
|
Display Screen |
5.7-inch TFT touch screen |
|
AC 220 V, ±10%, 50/60 Hz |
|
|
Fuses |
6 A |
|
Maximum Power Consumption |
300 W |
|
Memory Space |
Infinite (PC) |
|
Communication Methods |
Ethernet LAN, RJ45 |
|
External Trigger Input |
BNC, 5 V TTL |
|
CRO Trigger Output |
BNC, 5 V TTL |
|
Operation Control input |
BNC, 5 V TTL |
|
External Synchronous Input |
45-65 Hz, 20-500 V between dotted terminal & undotted terminal |
|
Failure Detection |
Display on LCD and interrupt the work at failure case |
|
Working Status Indication |
|
|
Grounding Connection Method |
Flat grounding cable |
|
Measurement Output Method |
Standard 1-meter coaxial cable |
|
Dimension |
35U rack cabinet, 600 mm (L)*800 mm(W)*1830 mm (H) |
|
Weight |
Approx. 450 kg |
|
Ambient Temperature |
15 ℃ - 35 ℃ |
|
Relative Humidity |
45% - 75% |
|
Atmospheric Pressure |
86 kPa - 106 kPa |
|
Technical Parameters |
|
|
In Compliance with |
IEC 61000-4-34 |
|
EUT Input Voltage |
Max. 690 V, 50/60 Hz |
|
EUT Current |
Max. 202 A cont. |
|
EUT Voltage/Current Measurement |
LCD display |
|
Withstand Impulse Current |
>1,000 A |
|
Protection |
Built-in over current, short circuit, over voltage, over heat, surge, burst, ESD protection; Optional external accessories for Line in, Line out port |
|
Interruption Level |
|
|
Dips Level |
0%, 40%, 70%, 80% |
|
Voltage Dips Mode |
Line to neutral & Line to line IEC61000-4-34: Figure3a, 3b |
|
0.3-9,999 cycles or 10 ms -99,999 ms |
|
|
Interval for Dips and Interruptions |
50 ms – 50,000 ms |
|
Test Counter |
1 – 9,999 |
|
Rise and Fall Time for Dips and Interruptions |
1-50 μs (standard resistance calibration) |
|
Sync |
0°~ 360°, 1°step or random mode |
|
Trigger Mode |
Manual, automatic, external trigger input |