General Parameters |
||
Display screen |
5.7 inch TFT touch screen 5.7 inch |
|
Power supply |
AC 110 V / 220 V±10%, 50/60 Hz ±5% (default AC 220 V 50 Hz) |
|
Fuse |
6 A |
|
Max. Power consumption |
200 W |
|
Memory space |
Infinite (PC) |
|
Communication mode |
Ethernet, RJ45 |
|
External control |
D-sub 25 p |
|
External trigger input |
BNC, 5 V TTL |
|
CRO trigger output |
BNC, 5 V TTL |
|
Input |
BNC, 5 V TTL |
|
Warning LED |
0-220 V AC 10 A/0-48 V DC 3 A Depending on external signal lamp |
|
Failure test |
Display on LCD and interrupt the work at failure case |
|
Working indicator |
LED, LCD front panel display |
|
Grounding connection |
Flat ground line |
|
Size |
19 inch/4U |
|
Weight |
Approx. 15 kg |
|
Temperature |
15 ℃ ~ 35 ℃ |
|
Air pressure |
86 kPa~106 kPa |
|
Impulse current test as per IEC 61643-1 and GB/T 18802.1 SURGGE module |
||
Model |
S6C 12 |
S6C 50 |
Test current |
1~120 A±10% |
30~500 A±10% |
Current waveform |
Front time: 8 μs ± 20 % Time to half value: 20 μs ± 20 % |
|
Charging voltage |
Setting current value*equivalent impedance + compensation voltage≦6000 V |
|
Source output impedance |
500 ohm(1-10 A) 50 ohm(11-120 A)) |
12 ohm |
Output polarity |
+/-/alternative |
|
Pulse cycle |
5 s ~ 99 s |
|
Surge current detection |
Front panel BNC output |
|
Surge current:10 A:1 V |
||
Counter |
1 - 10000 |
|
Trigger mode |
Automatic, manual and external |
Basic Equipment |
Simulator, user manual, test report, test cable, power cord and fuse |
Optional Accessories and Function |
1. HV differential probe, model:VCF-80, test voltage: 8 kV, attenuation ratio: 1000:1 2. HV current sensor, CM 0220M
3. PC control software CoreLab With PC installed WIN XP and WIN7, it can be easily operated to make the measurement, based on the customized test program. It can identify any device with automatic configuration. It can easily generate test reports.
|