General Description:
SEPN 3816T is the new generation of three-phase five-wire coupling/decoupling network for surge and burst test. It can be equipped with CCS600, CWS600 and EFT 500 or as customized requirements.
As Per:
IEC/EN 61000-4-4
IEC/EN 61000-4-5
ITU-T K.20
ITU-T K.21
ITU-T K.45
GB/T 17626.4
GB/T 17626.5
Features:
Relay protection after up to rating current
Temperature protection
Programmable set-up, save cost and time
|
General Data |
|
|
Operating Mains |
AC 110V/220V ±10%, 50/60Hz ±5% |
|
Max. power consumption |
100W |
|
Display |
LED |
|
Synchronization interface |
4mm banana plug |
|
EUT output |
4mm banana plug |
|
Burst input |
BNC |
|
Dimension |
19’’/6U |
|
Weight |
approx.35kg |
|
Temperature |
15℃-35℃ |
|
Humidity |
45%-75% |
|
Air pressure |
86kPa-106kPa |
|
Technical Data for decoupling network |
|
|
Model |
SEPN 3816T |
|
EUT rated voltage |
three-phase five-wire |
|
AC 380V±10%, 50/60Hz±5% DC110V±10% |
|
|
AC 220V±10%, 50/60Hz±5% DC110V±10% |
|
|
EUT Max.current |
AC16Acontinuous current |
|
DC16Acontinuous current |
|
|
300A impulse current |
|
|
Over current protection |
when current is more than32A |
|
Synchronization |
0°-359°, resolution 1° |
|
Residual voltage |
<15% |
|
Technical Data for surge coupling network |
|
|
Test voltage |
Max. 6kV |
|
Coupling waveform |
1.2/50μs; 8/20μs |
|
Coupling capacitance |
9μF,18μF |
|
Coupling resistance |
10Ω, 0Ω |
|
Attenuation ratio |
<10% |
|
Technical Data for burst coupling network |
|
|
Test voltage |
Max. 6kV |
|
Coupling waveform |
5/50ns |
|
Coupling capacitance |
33nF |
|
Resistance |
50Ω |
|
Frequency range |
0.1-1000 kHz |
|
Attenuation ratio |
<10% |
|
Accessories |
|
|
three-core power cord |
1 |
|
four-core cable |
1 |
|
coaxial cable |
1 |
|
connection wire, 25-core |
1 |
|
HV connection wire |
2 |
|
EUT connection wire |
5 |
|
Ground wire |
1 |
|
calibration certificate |
1 |
|
user manual |
1 |
|
test report |
1 |